Description
- Low and high temperatures (up to 1200 °C)
- Long term measurement at 660°C
- Useful for long term diagnostics (up to 30 years)
- Wide range of metallic, conductive and semi-conductive materials use
- Wireless access and on-line diagnostics
- Measuring during running conditions (heating, cooling, vibrating, deforming, …)
- Material degradation diagnostics by using of MDAPD Modified Direct and Alternating current Potential Drop technique (ACPD/DCPD/Corrected DCPD)
- Measurement applications (accuracy 1 ‰, resolution 0.01‰)
Applicable to:
- Insulated pipes and constructions (no need to remove insulation before measuring operation)
- At locations without near or open service access (up to 15 m measurement unit location distance and following wireless data transfer)
- Underground and high level locations (integrated measurement unit)
- Dangerous locations with radiation and chemical environments
- Explosive environments and areas with limited access
Uses / Applications
Measuring and monitoring of:
- Creep deformations – Permanent installation and diagnostics
- Crack initiation (fatigue) and propagation – Permanent installation and diagnostics
- Wall thickness reduction permanent installation and diagnostics
– corrosion losses
– material creep and deformation
- Crack depth – Diagnostics by using of contact method
- Crack on the opposite wall side – Permanent installation/contact method and diagnostics

Specification
- Four multiplexed channels
- 18-bit data acquisition, ±5V, 0.2% accuracy
- Measurement current 1A
- Communication interface – USB, RS232, RS485, Bluetooth, GSM, …
- 3.3” LCD display
- Up to 1000Hz measurement frequency, stability better than 0.01%
- Tunable gain 1 – 128x in 8 steps
- Tunable offset ±1000mV in 12 steps
- 12V, 1400mA power consumption
- Applicable to radiation enviroments